This guide provides step-by-step instructions for assembling a BX2 filter cube (U-MF2) for the Olympus BX2 and IX2 models (Chroma part number 91018), designed to precisely align optical filters in fluorescence microscopy systems. Proper assembly ensures optimal performance and system reliability.
For those new to handling optical filters, we recommend reviewing our Cleaning & Handling Instructions before assembling your filter cube.
Download the cube assembly instructions as an easy-to-use reference for the lab.
91018 BX2/IX2 filter cube
Exciter (x)
Ø25mm
Emitter (m)
Beamsplitter
38 x 26 x 1mm
A flat, metal spanner tool is used to tighten the retaining rings. This tool comes with the cube.
You also need a small Phillips-head screwdriver to separate the halves and loosen the clip that holds the beamsplitter down.
Separate the cube half by removing the two screws shown in the image.
Loosen the screws on the inside of the cube half with the clip that holds the beam splitter. It is not necessary to remove the clip; the filter will slide under it.
Mount the beam splitter by sliding it under the clip. The beam splitter should be mounted with the arrow facing up.
Tighten both screws on the clip to secure the beam splitter. Make sure the beam splitter is seated correctly before doing so, or you risk cracking the beam splitter.
Reassemble the cube halves. The exciter filter is inserted into the front port, with the arrow pointing toward the beam splitter.
Screw in the retaining ring with the spanner tool to secure the filter.
Insert the emitter filter into the port at the top of the cube, with the arrow pointing toward the beam splitter.
Finished Cube
Before installing the assembled cube into your optical system, confirm the following:
Verify that all filter arrows are aligned correctly. The beamsplitter arrow should face upward. The exciter and emitter filter arrows should point inward, toward the beam splitter.
Check that the registration tabs and holes between the cube halves are fully engaged. The two halves should meet cleanly, with no visible gaps or misalignment.
Ensure that the excitation and emission filters are properly installed. Filters should be fully inserted into their respective ports. Retaining rings should be threaded evenly to hold the filters securely, without tilt or cross-threading.
Confirm that the screws joining the cube halves are tightened. The cube should feel stable when handled, with no internal movement or looseness.
Inspect all optical surfaces for contamination and use cleanroom practices when handling filters. Refer to Chroma’s Cleaning and Handling Guidelines if cleaning is required.
If you encounter alignment issues, looseness, or other concerns during this inspection, disassemble and repeat the relevant steps. For additional support, please contact Chroma Technical Support.
If you don’t see what you are looking for, please contact us. We have a large inventory of filters and a knowledgeable staff ready to help you design and build a filter to meet your specifications.
10 Imtec LaneBellows Falls, VT 05101 USTel 800-824-7662Fax 802-428-2525sales@chroma.comMedia Queries: media@chroma.comTerms of Use Privacy Policy
Science, it's in our DNA!™
© 2026 Chroma Technology Corp. All rights reserved.
Please check the spelling or try searching with different keywords.
You can also explore the following sections for more information:
If you need further assistance, feel free to contact us:
We're here to help you find the optical filter solutions you need.